Figure 4-13: Clock Amplitude Initial Test Hookup - Sony Tektronix AWG2021 Service Manual

Arbitrary waveform generator
Table of Contents

Advertisement

Performance Tests
Check Clock Amplitude
4-32
Electrical
Auxiliary Outputs, Clock, Amplitude, on page 1-11.
Characteristic
Checked
Equipment
A 50 W coaxial cable
Required
One SMA BNC adapter
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a. Hook up oscilloscope: Connect the AWG2021 rear-panel CLOCK OUT
connector through a coaxial cable and SMA-BNC adapter to the
oscilloscope CH1 vertical input (see Figure 4–13).
AWG2021
AWG2021

Figure 4-13: Clock Amplitude Initial Test Hookup

b. Set oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Scale
. . . . . . . . . . . . . . . . . . . . .
Input impedance
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . .
Oscilloscope
TDS
540
CH1
DC
500 mV/div.
50 W
. . . . . . . . . . . .
500 ns/div.
AWG2021 Service Manual