Sony Tektronix AWG2021 Service Manual page 101

Arbitrary waveform generator
Table of Contents

Advertisement

AWG2021 Service Manual
Trigger
Source
. . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Select the waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to select the CLK_AMPL.WFM
file.
H Push ENTER.
3. Check clock amplitude accuracy: Check that the pulse displayed on the
oscilloscope has an amplitude of 0.7 V
4. End procedure: Disconnect the oscilloscope.
Performance Tests
CH1
DC
Positive
500 mV
Auto
to 1.3 V
.
p-p
p-p
4-33