7.2 Sample Racks

7.2 Sample Racks
Rack
code
X1
X2
1-10
57.1
178.6
11
59.0
176.6
12
57.2
178.5
13
68.5
167.5
14
68.5
167.5
15
66.1
169.6
16
66.1
169.6

7.2.1 Sample Racks Specification

a) Sample rack for thermostat (4 mL x 72)
Table 7-2 describes the sample racks that can be used with the
ChromasterUltra Rs 6270 Autosampler.
For using an optional sample rack, rack parameters should be set
so as to match that rack by Maintenance Software and CDS. (Refer
to Table 7-3)
Part No.
892-0240
Sample rack for
1.5 mL x 120 vials
892-0242
Sample rack for
4 mL x 72 vials
892-0244
Sample rack for
1 mL x 195 vials
892-0246
Sample rack for Microplate
(96 or 384 holes) x 2 plates
Table 7-3 Rack Parameters
Nx
Y1
Y2
10
0.0
148.5
34.0
8
15.4
149.8
48.0
13
8.6
150.0
49.0
12
89.1
152.1
42.0
12
2.8
65.8
42.0
24
87.0
154.5
42.0
24
0.5
68.0
42.0
i) Applicable product: ChromasterUltra RS 6270 Autosampler
ii) Temperature setup range: 1 – 45 °C (in increments of 1 °C)
iii) Temperature control range: [Ambient temperature -21 °C] to
[Ambient temperature + 25 °C] and within the
setup range (humidity 60 %, sample vial 4 mL)
iv) Temperature setup accuracy:
±3 °C
(Target temperature 4 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
sample vial (4 mL), without sample freezing)
Table 7-2 Sample Racks
Part Name
Pattern
Z
Ny
code
For 1.5ml×120 PCS (STD)
12
5
For 4ml×72 PCS
9
5
For 1ml×195 PCS
15
5
8
5
Back side of 96-holes microplate
8
5
Front side of 96-holes microplate
16
5
Back side of 384-holes microplate
16
5
Front side of 384-holes microplate
7 - 2
Remarks
Equipped as standard
-
-
Deep well plates are
unsuitable.
Application