Remove the syringe valve seal from the unit.
Syringe valve stator
(5)
In the reverse order of removal, attach a new syringe valve
seal, the stator ring, and the syringe valve stator. Then
tighten the stator setscrews. Secure these setscrews in turn
and by equal increments.
(6)
Push three tubes into the inner side (the syringe valve side)
as shown in Fig. 5-21, and connect to the syringe valve. Then
connect the other two tubes in the same manner.
Three tubes which put in the inner side.
The arm
Syringe valve seal
Figure 5-20 Removing the Syringe Valve Stator
Figure 5-21 Connection diagram of tubes
5 - 35
Unit
Stator setscrew
Push three tubes into the inner side.