Table 6-2 Autosampler Main Unit Errors (continued)
No.
Error Message
13
INJ TIMING ERROR
14
NO TUBE
15
STOPPED BY
EXTERNAL ERROR
16
ERROR-X
17
ERROR-Y
18
ERROR-Z
Possible Cause
1. Any pump other than the
Chromaster 6170 Pump is
combined for use with the
autosampler (when PASS
function is used.)
2. The flow rate specified for
the pump is 0 ml/min
(when PASS function us
used.)
1. No sample vials are set.
2. The clearance between
the top face of sample vial
and the detection lever is
inappropriate. Or the
detector lever is caught.
3. The electric system
malfunctions.
An error has occurred in other
unit connected.
1. The sample rack or
sample vial is blocking
movement.
2. There is an obstacle on
the moving route of arm X
(in left-right direction).
3. The electric system
malfunctions.
1. There is an obstacle on
the moving route of arm Y
(in front-rear direction).
2. The electric system
malfunctions.
1. There is an obstacle on
the moving route of needle
Z (vertically).
2. The electric system
malfunctions.
6 - 5
Remedy
1. Use it with the Chromaster
6170 Pump. Or set the PASS
function to "NO USE".
2. Set the flow rate correctly.
Alternatively, set the PASS
function to "NO USE".
1. Set sample vials.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
Eliminate the cause of the error to
release the error status.
1. Set the sample rack or sample
vial correctly.
2. Remove the obstacle.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. Remove the obstacle.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. Remove the obstacle.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.