4.2 Measuring Operation
(2)
Handling of Thermostat Rack
When the thermostat for the ChromasterUltra Rs 6270
Autosampler is used at high temperature and humidity place,
it may cause of dew condensation water in the sample rack.
Remove the sample rack to prevent splash of dew
condensation water to the instrument in the following
procedure.
(a)
Lift the handle of the sample rack (about 10 mm). The
locating pin of the sample rack is removed via the
locating hole of the rack holder. At this time, avoid lifting
the handle suddenly and highly. At this time, avoid lifting
the handle suddenly and highly. It may cause a
malfunction of the instrument, if dew condensation
water in the sample rack splashes to the drive part
behind the instrument.
(b)
Pull out the sample rack to the front side of the
instrument while keeping it in parallel position to the
rack holder.
4 - 6