5.3 Consumables and Spare Parts

5.3.7 Discharging Air Bubbles from Syringes

Plastic syringe
Figure 5-16 Connecting the Plastic Syringe (or Tube)
Items to be prepared:
Plastic syringe
A tube between a plastic syringe and a syringe valve
(1)
Detach the pipe from the port4 of the syringe valve.
Pipe
Syringe valve
Figure 5-15 Syringe Valve
(2)
Connect the provided plastic syringe to the port4.
(3)
Click PURGE for Syringe Purge (STROKE 10, SPEED 5) to
discharge wash fluid to the port4.
In the use of the plastic syringe, press the syringe with your
finger to apply proper pressure when wash fluid is discharged
through the port4. Thus, air bubbles can be removed with
ease (Figure 5-17.)
(4)
Click IDLE after you have completed eliminating air bubbles.
(5)
Attach the pipe on the port4 of the syringe valve.