String; Value - HP E1312A User Manual

Table of Contents

Advertisement

Example

:STRing?

Example

:VALue

Parameters
Comment
Example
:VALue?
Example
82 Multimeter Command Reference
Enter Calibration Information to Record the Next Calibration Date
CAL:STR 'Cal 4/4/YY, Due 10/4/YY' Enter a calibration message to record the cal
CALibration:STRing?
(or a null string " " if nothing is present).
Query the Calibration Message
CAL:STR?
enter statement
cal_value
CALibration:VALue <
signal used by the calibration procedure. See the HP E1312A and HP E1412A
Service Manual, Chapter 5 "Adjustments", for a more detailed description of the
multimeter's calibration/adjustment procedures.
Parameter Name
cal_value
<
>
does not affect the calibration value.
*RST
Enter the Known Value for the Calibration Source Signal
CAL:VAL 10.0
queries the present calibration value.
CALibration:VALue?
Query the Calibration Value
CAL:VAL?
enter statement
date of April 4 and next cal due date as
October 4 (YY = year of due date).
queries the calibration message and returns a quoted string
Query the calibration message.
Enter value into computer.
specifies the value of the known calibration
>
Parameter Type
Range of Values
numeric
See the service manual
Enter calibration value.
Query the calibration value.
Enter value into computer.
Default Units
none
Chapter 3