PERiod:APERture

Parameters
Comments
Example
PERiod:APERture?
Example
136 Multimeter Command Reference
[SENSe:]PERiod:APERture <
seconds. Values for time are rounded up to the nearest aperture time shown in the
following table.
Parameter Name
Parameter Type
time
<
>
sets the aperture time to 10ms.
MINimum
1 second.
The fastest aperture time available when autoranging is 100ms. In order to
specify an aperture time of 10ms, you must select a fixed range.
Setting the aperture time also sets the resolution. Aperture time of 0.01 sets
resolution at 4½-digits, 0.1 sets 5½-digits and 1 sets 6½-digits.
*RST
Condition: 0.1 (100ms)
Set the Aperture Time
PER:APER 1E-2
[SENSe:]PERiod:APERture? [MIN|MAX]
the output buffer:
The present aperture time in seconds if MIN or MAX is not specified.
The minimum aperture time available (10 ms) if MIN is specified.
The maximum aperture time available (100 ms) if MAX is specified.
Query the Aperture Time
PER:APER MIN
PER:APER?
enter statement
time
>|MIN|MAX
Range of Values
numeric
10ms|100ms|MIN|MAX
MAXimum
Aperture time is 10 ms.
returns one of the following numbers to
Aperture time is 10ms.
Query multimeter to return aperture time.
Enter value into computer.
sets the integration time in
Default Units
seconds
sets the aperture time to