Voltage[:Dc]:Aperture - HP E1312A User Manual

Table of Contents

Advertisement

VOLTage[:DC]:APERture

Parameters
Comments
Example
VOLTage[:DC]:APERture?
Example
Chapter 3
[SENSe:]VOLTage[:DC]:APERture <
for dc voltage measurements. Values are rounded up to the nearest aperture time
shown in the following table.
Parameter Name
Parameter Type
number
<
>
sets the aperture time to 0.333ms.
MIN
to 1.66667 seconds (60Hz) or 2 seconds (50Hz).
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz line
frequency) sets an integration time of 1 PLC. The corresponding resolution
depends on the function and range you select.
The
VOLT:APER
command overrides the results of previously executed
VOLT:NPLC
and
VOLT:RES
priority.
The greater the aperture time, the greater the normal mode rejection (and the
lower the reading rate).
Related Commands:
Condition:
*RST
VOLT:APER 0.20000
Set an Aperture Time of 16.7ms
VOLT:APER 16.7E-03
[SENSe:]VOLTage[:DC]:APERture? [MIN|MAX]
numbers to the output buffer.
The present aperture time in seconds if
The minimum aperture time available (.333ms) if
The maximum aperture time available (1.67s @ 60Hz; 2 s @ 50Hz) if
specified.
Query the Aperture Time
VOLT:APER 167E-03
VOLT:APER?
enter statement
number
>
Range of Values
numeric
0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
MAX
commands. The last command executed has
CALibration:LFRrequency
seconds (60Hz) or
VOLT:APER 0.166667
(50Hz)
Aperture time is 16.7ms.
MIN
Aperture time is 167ms.
Query multimeter to return aperture time.
Enter value into computer.
Multimeter Command Reference
sets the integration time in seconds
Default Units
seconds
sets the aperture time
returns one of the following
or
is not specified.
MAX
is specified.
MIN
is
MAX
147