VOLTage[:DC]:RESolution
Parameters
Comments
Example
VOLTage[:DC]:RESolution?
Chapter 3
[SENSe:]VOLTage[:DC]:RESolution <
voltage measurements.
Parameter Name
Parameter Type
resolution
<
>
•
MINimum
selects the best resolution (the smallest value) for the selected range.
selects the worst resolution (the largest value) for the selected range.
MAXimum
•
You must select a range using
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
•
If autoranging is required, set the resolution using the
or select a specific integration time using
•
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
executed has priority).
•
The
VOLT:DC:RES
CONFigure:VOLT:DC:RES
•
Related Commands:
•
Condition: Based on the
*RST
Change the Resolution
CONF:VOLT:DC 6.25,MAX
VOLT:DC:RANG 0.95
VOLT:DC:RES 3E-07
READ?
enter statement
[SENSe:]VOLTage[:DC]:RESolution? [MIN|MAX]
numbers to the output buffer.
•
The present resolution selected if
resolution values available on ranges set by the
returned.
•
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN
is specified.
•
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
is specified.
MAX
resolution
Range of Values
numeric
resolution
VOLTage:DC:RANGe
VOLTage:DC:NPLC
VOLTage:DC:NPLC
commands (the last command
command overrides the resolution setting from a previous
command.
,
CONFigure
VOLTage:AC:NPLC
values for the
*RST
Function: DC volts; range selected: 10V;
MAX resolution.
Range selected: 1V; MAX resolution.
Set resolution to 0.3µV.
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
Enter readings into computer.
or
MIN
MAX
Multimeter Command Reference
selects the resolution for DC
>
Default Units
volts
|MIN|MAX
before specifying
MIN
or
MAX
parameters
.
command.
VOLTage:NPLC
returns one of the following
is not specified. Only the
command are
RANGe
151