:VOLTage:AC?

Parameters
Comments
Example
108 Multimeter Command Reference
MEASure:VOLTage:AC? [<
[,<
resolution
>|MIN|MAX|DEF]]
allows you to specify the range and resolution (see range versus resolution table at
start of chapter).
Parameter Name
Parameter Type
range
<
>
resolution
<
>
To select a standard measurement range, specify range as the input signal's
maximum expected voltage. The multimeter then selects the correct range to
accept the input.
The
or
AUTO
DEFault
The
DEF
option for the resolution parameter defaults the integration time to
10 PLC.
The
and
MIN
MAX
range:
For range:
= 0.1V;
MIN
For resolution:
MIN
selected range.
MAX
selected range. See Table 3-4 on page
each range.
To select autorange, specify
range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
To specify a
MIN
or
MEAS:VOLT:AC? DEF
the
or
resolution from being interpreted as a range setting and the
MIN
MAX
resulting command becomes
MEAS:VOLT:AC? DEF,MAX
Making AC Voltage Measurements
MEAS:VOLT:AC? 0.54,MAX
range
>|MIN|MAX|DEF|AUTO
selects the AC-coupled RMS voltage function and
Range of Values
numeric
0.1V|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
numeric
resolution
option for the range parameter enables autorange.
parameters select the minimum or maximum values for
= 300V.
MAX
selects the best resolution (the smallest value) for the
selects the worst resolution (the largest value) for the
71
DEF
for range or do not specify a value for the
MAX
resolution while autoranging, you must specify
(you cannot omit the range parameter). This prevents
MEAS:VOLT:AC? DEF,MIN
.
Function: AC volts; range selected: 1V;
MAX resolution: 100µV.
Default Units
|MIN|MAX|DEF
for valid resolution choices for
or
volts
volts
Chapter 3