Default Delays

50 HP E1312A/E1412A Multimeter Application Information
If you do not specify a trigger delay, the multimeter automatically
determines a delay time (default delay) based on the present measurement
function, range, resolution, integration time and AC filter bandwidth setting.
The delay time is actually the settling time required before measurements to
ensure measurement accuracy. The default delay time is automatically
updated whenever you change the function or range. Once you specify a
delay time value, however, the value does not change until you specify
another value, reset the multimeter or do a
table below shows the default trigger delay times for all functions. This
delay will occur before each measurement (see the trigger system diagram
in Figure 2-1 on page 45).
NOTE: You can specify a shorter delay time than the default values shown.
However, the shorter settling time may not produce accurate measurements.
Table 2-7. Default Trigger Delays
Default Trigger Delays for DC Voltage and DC Current (all ranges):
Integration Time
NPLC ≥1
NPLC <1
Default Trigger Delays for 2-Wire and 4-Wire Resistance:
Range
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
Default Trigger Delays for AC Voltage and AC Current (all ranges):
AC Filter
3Hz - 300kHz filter
20Hz - 300kHz filter
200Hz - 300kHz filter
Default Trigger Delay for Frequency and Period:
or
CONF
MEAS
Trigger Delay
1.5ms
1.0ms
Trigger Delay
(For NPLC ≥1)
1.5ms
1.5ms
1.5ms
1.5ms
1.5ms
100ms
100ms
Trigger Delay
7.0sec
1.0sec
600ms
1.0s
command. The
Trigger Delay
(For NPLC <1)
1.0ms
1.0ms
1.0ms
1.0ms
10ms
100ms
100ms
Chapter 2