Measurement Speed And Accuracy Trade-Offs; Hp E1312A/E1412A Special Function And Range Commands (Non-Scpi ) - HP E1312A User Manual

Table of Contents

Advertisement

Measurement Speed and Accuracy Trade-offs

HP E1312A/E1412A Special Function and Range
Commands (Non-SCPI )
FUNCTION
R1
F1 (DCV)
0.1V
F2 (DCI)
0.01A
F3 (RES)
100Ω
F4 (FRES)
100Ω
Appendix C
The HP E1312A and HP E1412A Multimeters were designed so the default
mode of operation will deliver high accuracy readings with a minimum of
programming effort. However, many applications require high-speed
measurements. This appendix discusses two topics:
1. special non-SCPI function (
,
,
and
R4
R5
R6
and
2. how to increase measurement speed where reduced reading accuracy
is acceptable.
The HP E1312A and HP E1412A Multimeter have special function and
range commands for
resistance (
FRES
) shown in the following table. ACV, ACI, frequency or
period functions are not supported with these special commands.
Special Function and Range Commands
R2
R3
1V
10V
0.1A
1A
1KΩ
10KΩ
1KΩ
10KΩ
These special commands act like a
or range. The range command acts only on the current function. For
example, if the current function is
range command
changes the
R2
,
or
ranges. To also change the 2-wire resistance range to the
DCI
RES
FRES
R2 setting, you must send the commands
current function from
returns the function to
F1
to the
F3
command.
The table on the following page shows equivalent
the special commands.
Measurement Speed and Accuracy Trade-offs
F1
,
F2
,
F3
and
) commands used to speed up measurement setup
R7
DCV
,
DCI
, 2-wire resistance (
RANGE
R4
R5
100V
300V
3A
WILL RETURN "Out of range" ERROR
100KΩ
1MΩ
100KΩ
1MΩ
command to change a function
[SENSe:]
DCV
and its range is 10V, sending the
range to 1V but does not affect the
DCV
F3
to
then
sets the range to 1KΩ. Sending
DCV
RES
R2
and to the range and state it was last set prior
DCV
Appendix C
F4
) and range (
R1
,
R2
RES
) and four-wire
R6
R7
INVALID RANGES
10MΩ
100MΩ
10MΩ
100MΩ
and
R2
. First
F3
changes the
[SENSe:]
commands for
,
R3
,
193