HP E1312A/E1412A Resolution Using Special Functions
and Ranges

Resolution Example

Appendix C
Resolution remains a function of the NPLC parameter set at the time a
special function or range is used. The NPLC setting is fixed throughout use
of the special functions and ranges unless you change the setting with the
[SENSe:]<
function
CONFigure command using a resolution that changes the NPLC setting. The
resolution will track the NPLC setting as shown in Tables 3-1, 3-2, and 3-3
beginning on page 70.
The following table shows range and NPLC settings for power-on and after
a module reset. Changing a range within one function does not place other
functions at that range setting. Each function operates independently.
Special Function
F1
F2
F3
F4
Note
Refer to Tables 3-1, 3-2 and 3-3 in Chapter 3, to determine what resolution
will result following a special function or special range change. The NPLC
setting remains fixed for each function during execution of the special
function and range commands (e.g., may differ from function to function).
Assume the power-on state where the multimeter function is DC Voltage,
10V range, with an NPLC setting of 10 PLCs providing 10µV resolution
(see Table 3-1 on page 70). Use the special range command R5 to change
the DC Voltage range to 300V. The NPLC setting remains at 10 PLCs
providing a resolution of 1mV (see Table 3-1 on page 70).
Use the special function command
Resistance. The range goes to the resistance power-on state (1k
Resistance, NPLC = 10); it does not change with the previous
command. NPLC remains at 10 PLCs providing resolution of 1m
The special range commands do not affect other functions except in the
(
RES
) and
F4
same range change on
>:NPLC
command or configure the multimeter with the
HP E1312A/E1412
Power-on State
10V (R3), NPLC = 10
1A (R3), NPLC = 10
1kΩ (R2), NPLC = 10
1kΩ (R2), NPLC = 10
F3
(
FRES
) function changes. Range changes on
and vice versa.
F4
Measurement Speed and Accuracy Trade-offs
HP E1312A/E1412
*RST State
300V (R5), NPLC = 10
1A (R3), NPLC = 10
1kΩ (R2), NPLC = 10
1kΩ (R2), NPLC = 10
to change the function to 2-Wire
F3
for 2-Wire
DCV R5
.
F3
cause the
195