Querying the
Delay Time
The Sample Count
Example: Setting the
Sample Count
Checking the
Sample Count
Chapter 2
The
TRIGger:DELay? [MINimum|MAXimum]
following numbers to the output buffer:
•
The present trigger delay (1µs through 3600 seconds) if
not specified.
•
The minimum trigger delay available (1µs) if
•
The maximum trigger delay available (3600 seconds) if
specified.
The
SAMPle:COUNt <
number
readings per trigger. The number parameter sets the number of readings to
a value between 1 and 50,000.
Substituting
MIN
for the number parameter sets the number of readings per
trigger to 1. Substituting
readings per trigger to 50,000.
In the following example, 10 DC voltage measurements are made when the
multimeter's external trigger BNC connector is pulsed low. After the
10 readings are taken, the multimeter returns to the idle state.
dimension array
CONF:VOLT:DC
TRIG:SOUR EXT
SAMP:COUN 10
READ?
timeout may occur
enter statement
The
SAMPle:COUNt? [MINimum|MAXimum]
following numbers to the output buffer:
•
The present sample count (1 through 50,000) if neither
is specified.
•
The minimum sample count available (1) if
•
The maximum sample count available (50,000) if
HP E1312A/E1412A Multimeter Application Information
command returns one of the
>
command designates the number of
for the number parameter sets the number of
MAX
Dimension computer array.
Function: DC voltage.
Trigger source is external BNC on multimeter front
panel.
Specify 10 readings per trigger.
Place multimeter in wait-for-trigger state; make
measurements when external trigger is received; send
readings to output buffer.
May require INIT, monitor the status byte for
completion (standard event bit 0), FETC? to transfer
readings to the output buffer (vs. READ?).
Enter readings into computer.
command returns one of the
MIN
or
MAX
MIN
is specified.
MAX
is
nor
MIN
MAX
MIN
is specified.
MAX
is specified.
is
51