RESistance:NPLC

Parameters
Comments
Example
RESistance:NPLC?
Example
140 Multimeter Command Reference
[SENSe:]RESistance:NPLCycles <
line cycles (PLCs). The NPLC is set to a value from the range of values that can
accommodate the
<
number
to 100.
Parameter Name
Parameter Type
number
<
>
selects 0.02 PLCs.
MINimum
integration time in power line cycles (PLCs) also sets the integration time and
the resolution. For example, 10 PLCs sets an aperture time of 167ms (60Hz
line frequency) or 200ms (50Hz). The corresponding resolution depends on
the function and range you select.
The
command overrides the results of a previously executed
RES:NPLC
RESistance:RESolution
command executed has priority).
The greater the number of PLCs, the greater the normal mode rejection (and
the lower the reading rate).
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode rejection
of 50Hz or 60Hz power line related noise.
*RST
Condition: 10 PLC
Set the Integration Time in PLCs
RES:NPLC 100
[SENSe:]RESistance:NPLC? [MIN|MAX]
the output buffer:
The present integration time in PLCs if
specified.
The minimum integration time available (0.02) if
The maximum integration time available (100) if
Query the Integration Time
RES:NPLC 100
RES:NPLC?
enter statement
number
sets the integration time in power
>
>
you specify. For example, specifying
Range of Values
numeric
0.02|0.2|1|10|100|MIN|MAX
selects 100 PLCs. Setting the
MAXimum
or
RESistance:APERture
Integration time is 100 PLCs.
returns one of the following numbers to
MINimum
Integration time is 100 PLCs.
Query multimeter to return integration time.
Enter value into computer.
11
sets the NPLC
Default Units
PLCs
command (the last
or
is not
MAXimum
MIN
is specified.
is specified.
MAX
Chapter 3