Chapter 6: Diagnostics; Introduction; Displaying Diagnostic Parameters - GE AquaTrans AT868 User Manual

Ultrasonic flow transmitter for water (1- & 2-channel)
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

Introduction

Displaying Diagnostic
Parameters
Accessing the Display
Submenu
Configuring the Display
Diagnostics
This chapter explains how to troubleshoot the Model AT868 if
problems arise with the electronics enclosure, the flowcell, or the
transducers. Indications of a possible problem include:
display of an error message on the LCD screen
erratic flow readings
readings of doubtful accuracy (i.e. readings that are not consistent
with readings from another flow measuring device connected to
the same process).
If any of the above conditions occurs, proceed with the instructions
presented in this chapter.
The Model AT868 has built-in Diagnostic Parameters to aid in the
troubleshooting of flowcell, transducer and/or electrical problems. To
access these parameters, do the following (refer to Figure A-2 on
page A-2 in Appendix A, Menu Maps):
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until GLOBL appears and press [ENTER].
4. Press [ ] until I/O appears and press [ENTER].
5. Press [ ] until LCD appears and press [ENTER].
1. At # of LCD PARAMS, press [ ] to select the desired number of
parameters to be sequentially displayed and press [ENTER].
For a 1-channel/path AT868, proceed to Step 3. For a 2-channel/path
meter, proceed to Step 2.
Note: If you select OFF, the meter returns to the Global I/O window.
Go to Procedure Options on page 6-3.
2. Press [ ] to select the desired channel/path option and press
[ENTER].
3. At Measurement Name, press [ ] to select the desired parameter
(refer to Table 6-1 on the next page for a list of available
diagnostic parameters) and press [ENTER].
September 2004
6-1