GE AquaTrans AT868 User Manual page 57

Ultrasonic flow transmitter for water (1- & 2-channel)
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

September 2004
Setting Up a Totalizer
Output
Procedure Options
2-32
1. At Output B or D, select OFF or TTLZR and press [ENTER].
2. Do one of the following:
If you selected OFF, the meter returns to the Global I/O window.
Go to Procedure Options on page 2-32.
If you are using a 1-channel meter, go to Step 4.
If you are using a 2-channel/path meter, go to Step 3.
3. At Channel, press [ ] to select the desired channel/path and press
[ENTER].
4. Press [ ] to select the desired measurement parameter and press
[ENTER].
5. Enter a value for the minimum pulse on time (between 50 µsec
and 500,000 µsec) for the frequency of the totalizer pulses and
press [ENTER]. Press [
[ ]
location and press
number.
Note: A complete pulse consists of equal amounts of ON and OFF
times. Choose a value that is compatible with the counter to be
used.
6. At UNITS/PULSE enter a value for the number of measurement
units represented by each pulse and press [ENTER].
After completing the above steps, the user program returns to the
Global I/O window. Do one of the following:
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] four times.
]
and [ ] to move the cursor to desired
[ ]
and
to increase or decrease the
Programming Site Data