KV Input Selection

Programming Site Data
]
1. Press [
and [ ] to enter a static kinematic viscosity or a table
of values and press [ENTER].
2. Do one of the following:
If you selected Table, enter the Calibration Factor by using the
arrow keys. Press [
location and press
number and press [ENTER].
Note: The values in the KV/SS table can be edited as described in
Entering KV/SS Table on page 2-18.
If you selected STATC, the AT868 will select and automatically
display the Kinematic Viscosity (see note below). If you want
to enter a different number use the [
cursor to desired location and press
or decrease the number and press [ENTER]. Then enter the
Calibration Factor in a similar manner and press [ENTER].
Note: If you select OTHER as the fluid type, you must enter the
Kinematic Viscosity. The AT868 will only approximate the
Kinematic Viscosity if you select one of the available fluid
options other than OTHER (i.e. water, oil, etc.).
3. Proceed to one of the following sections:
For Clamp-on Transducers - proceed to Number of Traverses
and Transducer Spacing on page 2-12.
For Wetted Transducers - The meter returns to the Channel
PROGRAM window. At this point, you can continue
programming in this submenu as desired. To leave the Channel
PROGRAM window, press [ESC] three times. To complete
setting up the meter, you must enter data in the GLOBL-SYSTM
submenu as described on page 2-22.
]
and [ ] to move the cursor to desired
[ ] and [ ]
to increase or decrease the
]
and [ ] to move the
[ ] and [ ]
September 2004
to increase
2-11