Setting Up the External
Totalizer Reset Switch
Programming Site Data
The GATE OPTION enables you to configure an external reset switch
that can reset totals by activating a switch that is remotely located
from the electronics. To connect an external reset switch, refer to
Wiring the External Totalizer Reset Switch on page 1-12.
IMPORTANT:
The hardware for the switch is user-supplied. GE
Infrastructure Sensing supplies only the software to
configure the switch and test points to connect the
switch.
6. At the GATE OPTION prompt, press [ ] to select one of the
following and press [ENTER].
RESET - the meter resets totals to zero for both channels and
•
then immediately begins to accumulate again.
HOLD - the meter holds totals at the current point for both
•
channels. The meter will not accumulate as long as the switch
remains closed. This option is typically used for calibration.
7. Do one of the following:
•
If MASS FLOW is ON and both channels are active (for
2-channel units only), proceed to Selecting Mass Flow Units on
page 2-24. Both channels must be active
•
If MASS FLOW is OFF and/or one channel is active, the meter
returns to the Global PROGRAM window. Go to Procedure
Options on page 2-24.
Note: To activate mass flow, refer to page 2-20.
September 2004
2-23