Setting Up Analog And Totalizer/Frequency Outputs - Optn - GE AquaTrans AT868 User Manual

Ultrasonic flow transmitter for water (1- & 2-channel)
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

September 2004
Setting Up Analog and
Totalizer/Frequency
Outputs - OPTN
Accessing Analog Outputs
Setting Up the Analog
Output Scale
2-28
The Model AT868 has one built-in analog output and one totalizer/
frequency output for each channel which must be set up. To
accomplish this, refer to one of the following sections:
Analog Output - refer to the section below.
Totalizer/Frequency Output - refer to page 2-30.
Output A and C are analog outputs that can be configured
independently. Typically Output A is used for Channel 1 and Output
C is used for Channel 2; however, analog outputs can be configured to
measure either channel.
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until GLOBL appears and press [ENTER].
4. Press [ ] until I/O appears and press [ENTER].
5. Press [ ] until OPTN appears and press [ENTER].
6. Press [ ] to select A or C and press [ENTER].
1. Press [ ] to select the desired output scale and press [ENTER].
2. Do one of the following:
If you selected OFF, the meter returns to the Global I/O window.
Go to Procedure Options on page 2-29.
If you are using a 1-channel meter, go to Step 4.
If you are using a 2-channel/path meter, go to Step 3.
3. At Channel, press [ ] to select the desired channel/path and press
[ENTER].
4. Press [ ] to select the desired measurement parameter and press
[ENTER].
Note: The measurement units that appear in these prompts are those
GLOBL-SYSTM
selected in the
menu earlier in this section.
Programming Site Data