Siemens FUJI-Reader CR-IR347 Service Manual page 451

Hide thumbs Also See for FUJI-Reader CR-IR347:
Table of Contents

Advertisement

I SNS08C board
TP1
F2
CN4
F1
CN12
CN10
CN11
CN15
CN16
CN25
CN8
G Fuse Ratings
Location Name
F1
F2
F3
G Checking the Fuses
Measure the resistance value across the fuse. If the measured resistance value is 100 Ω or
greater, it is judged that the fuse has been blown.
Fuse
SUPPLEMENTARY NOTES
• For the tester, its probe resistance (resistance value at the time of probe shorting) should
be measured before measurement, and compensation should be made with respect to the
measured value of the fuse.
• You cannot accurately determine whether a fuse is blown or not while the associated board
is mounted in the machine. However, if the resistance value measured across the fuse
with the board installed in the machine is 1 Ω or greater, it is highly probable that the fuse
is blown. In such an instance, demount the board and check whether the fuse is actually
blown.
009-058-03
08.30.2002 FM3476
1
9
2
10
F3
CN2
CN22
CN23
CN13
CN7
CN9
Rated voltage
125 V
125 V
125 V
Tester
FR7H3823.EPS
CR-IR347 Service Manual
1
9
2
10
CN3
T1
T2
CN18
CN5
CN6
CN14
Rated amperage
5 A
2 A
1 A
MC - 424
CN1
FR7H3844.EPS
MC - 424