I Installation procedure
To accomplish installation, reverse the removal procedure.
I Check/adjustment procedure
• Make sure that the anti-static member is brought into contact with the roller.
• Check that the long timing belt is properly threaded through its path.
• After the long timing belt is replaced with a new one or the tensioner is removed and
reinstalled, remove the erasure conveyor and adjust the belt tension with the tensioner
screw.
"6.2 Erasure Conveyor"
5±1 mm
009-058-02
009-058-00
08.30.2001 FM3142
10.20.2000 FM2732 (2)
Scales
CR-IR347 Service Manual
Push-pull gauge
5±0.5 N
[Adjust] Tensioner
Timing belt
FRONT
MC - 162
FR7H3526.EPS
MC - 162