Siemens FUJI-Reader CR-IR347 Service Manual page 264

Hide thumbs Also See for FUJI-Reader CR-IR347:
Table of Contents

Advertisement

[5] Correction result verification
(1) Expose the IP.
❍ ST
IP used:
X-ray tube target:
Tube voltage:
Dose:
❍ HR
IP used:
X-ray tube target/filter:
Tube voltage:
Dose:
(2) Using the test menu, record and output an image.
❍ ST
"
❍ HR
"
"SENSITIVITY2" (for versions other than Japanese and U.S.)
N NOTE N
Generate an image output when 10 minutes elapse after IP exposure.
(3) Run the following checks on the obtained output film:
❍ Check for image irregularities.
❍ Check that the S value prevailing on the film is within the following range:
<ST>
<HR>
(4) Terminate the shading correction/sensitivity correction process.
009-058-03
08.30.2002 FM3476
" (for Japanese version)/"SENSITIVITY" (for versions other than Japanese)
" (for Japanese version)/"SensitivityHQ" (for U.S. version)/
140 ≤ S value ≤ 260
84 ≤ S value ≤ 156
CR-IR347 Service Manual
14" x 17" (14" x 14" may also be used if
14" x 17" is not available.)
Tungsten
80 kV
1 mR
IP used: 24cm x 30cm mammo (18cm x
24cm mammo may also be used if 24cm x
30cm mammo is not available.)
Molybdenum/molybdenum
25 kV
20 mR
MC - 257
MC - 257