I Installation procedure
N NOTE N
When installing the LD assembly, properly position it by pressing it against the three
positioning brackets.
Accomplish installation by reversing the removal procedure.
I Check procedure
CAUTION
Before removing a cover, be sure to turn OFF the high-voltage switch. Removing a cover
with the machine turned ON causes photomultiplier damage.
(1) Turn ON the machine.
(2) Verify that the machine starts up normally.
(3) Check for image problems.
"9.3 Checking for Image Problems"
009-058-00
10.20.2000 FM2732 (2)
DETAIL A
FRONT
Positioning brackets
Scanning
optics unit
Scanner optics unit
CR-IR347 Service Manual