(3) Remove the grip release wires.
DETAIL A
B3x6
Wire
009-058-00
10.20.2000 FM2732 (2)
Triangular cam
[Remove] E4 (x3)
A
Guard
#1
[Remove] B3x6 (x3)
B
Bracket
Spacer
Bracket
#2
[Remove] BR4x35
Triangular cam
Na3
(Loktite)
Shaft
CR-IR347 Service Manual
#3
Pulley
A
Shaft
X
C
DETAIL B
Na3 (Loktite)
Triangular cam
B3x6
Shaft
Wire
MC - 236
DETAIL C
Wire (long)
Wire (short)
Wire (medium)
E3
FR7H3136.EPS