Siemens FUJI-Reader CR-IR347 Service Manual page 43

Hide thumbs Also See for FUJI-Reader CR-IR347:
Table of Contents

Advertisement

I Installation procedure
N NOTE N
• When installing the claw guide assembly, adjust the degree of guide plate parallelism
with the mounting screws (BR4x8, x2) loosened.
For parallelism adjustment purposes, apply a 300 mm scale to the guide plate trailing
end (toward the IP removal unit) and measure the gap between the scale and guide
plate.
009-058-00
10.20.2000 FM2732 (2)
0.5 mm or greater gap = NG
Gap smaller than 0.5 mm = OK
300 mm scale
Less than 0.5 mm
Guide plate
CR-IR347 Service Manual
300 mm scale
MC - 36
FR7H3320.EPS