I Fuse Ratings
Location Name
F1
F2
F3
F5
F6
I Checking the Fuses
Measure the resistance value across the fuse. If the measured resistance value is 100 Ω or
greater, it is judged that the fuse has been blown.
Fuse
SUPPLEMENTARY NOTES
• You cannot accurately determine whether a fuse is blown or not while the associated board
is mounted in the machine. However, if the resistance value measured across the fuse
with the board installed in the machine is 1 Ω or greater, it is highly probable that the fuse
is blown. In such an instance, demount the board and check whether the fuse is actually
blown.
• For the tester, its probe resistance (resistance value at the time of probe shorting) should
be measured before measurement, and compensation should be made with respect to the
measured value of the fuse.
I Action to be Taken When the Fuse Has Been Blown
Replace the fuses.
009-058-03
08.30.2002 FM3476
Rated voltage
125 V
125 V
125 V
125 V
125 V
FR7H3834.EPS
CR-IR347 Service Manual
Rated amperage
2 A
2 A
2 A
4 A
5 A
MC - 405