G After the light-collecting guide assembly is replaced with a new one
(1) Rotate the lever to secure the light-collecting guide assembly (for back surface). Note
that the assembly is not to be screwed down (BR4x8).
Steps (1) through (3) under "G Removing/installing the light-collecting guide
assembly"
(2) Reposition the bracket as needed and then secure it.
009-058-00
10.20.2000 FM2732 (2)
Subscanning unit
Bracket
[Loosen] Positioning screw
#4
[Contact] Bracket
[Fasten] Positioning screw
CR-IR347 Service Manual
Light-collecting guide assembly
(for back surface)
#3
#5
#6
[Contact] Bracket
MC - 317
#1
[Loosely attach]
Adjustment screw
Side plate
#2
[Loosen]
Adjustment screw
#9
[Tighten] BR4x8
#7/#8
[Contact/Tighten]
Adjustment screw
FR7H3634.EPS
MC - 317