GE eBike III basic Series Service Manual page 97

Electric bike with and without nibp
Table of Contents

Advertisement

e
Bpu (B
rror Code BIts
Bit
Error
0
not used
1
cuff loose
2
Micro position
3
Leakage
4
Errror Calibration
5
Error offset
6
not used
7
not used
8
Safety state mode
9
Setup not ready
10
CAN zykl. Timeout
2018111-336 Rev. B
)
lood pressure Module
Reaction
Module is switched off
Module is switched off
Module is switched off
Module is switched off
Module is switched off
eBike III basic / comfort
Display / Info
Troubleshooting
Warning
- check correct placement of cuff
at patient (tight)
- check cuff
Warning
- check correct placement of
microphone (see User Manual)
- check microphone in cuff
- replace cuff
Warning
- check cuff
- check hoses, hose connections
- check compensation container
- replace pneumatic unit
Internal error
- perform NIBP test
- perform NIBP calibration
- replace BPU PCB
Internal error
- perform NIBP test
- perform NIBP calibration
- replace BPU PCB
Internal error
- restart eBike (OFF / ON)
- if error persists:
replace BPU
Internal error
- restart eBike (OFF / ON)
- if error persists:
check connection LIU <-> BPU
- replace BPU
Internal error
- restart eBike (OFF / ON)
- if error persists:
check cable DTU <-> LIU
check connection LIU <-> BPU
- replace BPU
- 97 -

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ebike iii comfort series

Table of Contents