Nibp Test - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

All the components are connected by tubes (one pressure
system) and connected to the cuff connector of the
ergometer control terminal (use cuff adaptor).
nIBp t
est
This function is used to check the internal pressure system
(pressure sensor, closeness, leakage etc.).
Select "NIBP Test" in the service menu and confirm by
pressing NIBP.
Use the hand pump to apply a fixed pressure to the
system.
The actual pressure value is displayed.
Compare the displayed ergometer value with the value of
the manometer.
The maximum allowed deviation is +/- 3mmHg.
Select EXIT (and confirm by pressing NIBP) to return to the
service menu.
- 76 -
compensating container
500 ml
c
alibration Setup

NIBP Test

NIBP pressure
CLOSE VALVE
EXIT
t
eSt of the internal
If the applied pressure reaches 320 mmHg, the internal
safety valve is opened
Select „Close valve" (and confirm by pressing NIBP) to
continue with the NIBP-Test
eBike III basic / comfort
Control unit
M
Connecting tubes
manometer
nibp
123 mmHg
nibp
preSSure SyStem
Note
2018111-336 Rev. B
hand pump