Software Update - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

s
oftware update
To load a new version of the internal software into the
eBike III ergometers; a PC or Laptop with USB or a serial
port (RS-232) is needed.
To use the ergometer USB port, a special driver has to be
installed on the PC (virtual COM driver).
Copy the complete software package onto the PC and
uncompress it.
Connect the PC or Laptop with a serial cable or USB cable
to the egometer (Port 1 oder USB).
The software update is initiated in the service menu.
To activate the service menu, the ergometer has to be
powered off and on again.
As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously:
To prevent settings from being changed in error, the
service menu is "protected" with a password.
Select "999" with the arrow keys and confirm by pressing
NIBP.
The complete handling of the menues is done using the
arrow keys (saddle height) to position the bar cursor and
the NIBP key (near the display) to confirm the selection.
- 78 -
Always use the latest software version for the appropriate
ergometer
Use only the PC update program included in the software
package
S
-
elf
teSt Screen
p
aSSword for Service menu
eBike III basic / comfort
Note
eBike III
Selftest running
Service Code
999
2018111-336 Rev. B

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ebike iii comfort series

Table of Contents