6. Write down the assigned COM port; this port number
has to be used in the CardioSoft/CS settings.
7. Close all open windows.
C
CoM
Hange
port assIgnMent
1. Right click "Silicon Labs CP 210x USB to UART Bridge
(COMx)"
2. Select "Settings"
3. Select tab "Port Settings".
4. Click "Advanced".
5. For CardioSoft/CS V6.71 and below:
Click the "COM Port Number" and select an unused
COM port between COM1 and COM4.
6. For CardioSoft/CS V6.73 and above:
Click onto "COM Port Number" and select an unused
COM port between COM1 and COM32.
7. Confirm by clicking OK.
8. Close all open windows.
B
III
e
Ike
software settIng
To activate the service menu, the ergometer has to be
powered off and on again.
As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously:
To prevent settings from being changed in error, the
service menu is "protected" with a password.
Select "999" with the arrow keys and confirm by pressing
NIBP.
2018111-336 Rev. B
eBike III basic / comfort
Service Code
999
- 85 -