Test Procedure "Rpm / Load - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

t
"rpM / l
est proCedure
Unless regulatory requirements of the country define a different test procedure or test intervall, it is recommended to perform the
test "RPM / load" at least every 2 years
An external calibration device for dynamic load testing is needed to perfom the following test procedures
("ergoline ergotest 550" - order direct from Ergoline P/N 705890) or equivalent device
Please refer to the operator manual of the used calibration device
t
est preparatIon
1. Dismount BOTH (!) pedal cranks of the eBike III - see "Dismount pedal cranks" under section „Dismount pedal cranks"
on page 41.
2. Follow the operator manual from the manufacture of calibration device being used to connect the calibration device
with the axis of the eBike III.
3. If supported by the calibration device, connect the eBike III interface (RS-232 or USB) with the calibration device.
4. Connect the calibration device to the PC / notebook setup containing the software and configuration to work with the
device."
5. Start the test software of the calibration device.
rpM
test protoCol
1. Apply a load of 100 W to the eBike III
(-either by command of the test software or manually on the eBike III (see „Manual Load" on page 66).
2. Perform the following RPM tests step by step. Note the results in the table or use the printed protocol of the
calibration device to document the test.
Step
Target RPM
#1
30
#2
40
#3
50
#4
100
#5
120
2018111-336 Rev. B
"
oad
Note
Note
Actual RPM
eBike III basic / comfort
Deviation max.
± 1 rpm
± 1 rpm
± 1 rpm
± 1 rpm
± 1 rpm
Pass / Fail
- 15 -