nIBp
Blood pressure unIt
a
sseMBly
The NIBP module consists of several components.
The complete measurement electronics, the deflating valve
and the power drivers for the pump are placed on the
BPU board (see picture).
The pump and the measuring valve are placed on a
separate sound and vibration absorbing foam carrier.
When replacing the measuring valve, the silicon tube has
to be connected with the upper hose nipple of the valve
and the cable bended to a loop. The carefully slip the valve
back into the foam carrier.
- 32 -
bpu m
eaSuring electronic board
bpu m
eaSuring electronic board
1
microphone connector (J 9)
2
measuring valve connector (J 6)
3
pump connector (J 4)
f
oam carrier for meaSuring valve
(
with compenSating container
r
eplacing the valve
eBike III basic / comfort
1
2
3
1
2
3
(1)
(2)
and pump
3)