Honeywell SmartLine ST 700 User Manual page 74

Pressure transmitters
Hide thumbs Also See for SmartLine ST 700:
Table of Contents

Advertisement

Barometric Correct
[27BARCOR]
Calibrate High Point
[28CAL HI]
Calibrate Low Point
[29CAL LO]
Calibrate Adjust Ratio
[30CALRTO]
Page 58
One point field offset
Calibration Method.
######. ##
######. ##
#. ##
ST 700 SmartLine Pressure Transmitters User's Manual
Available for AP models only.
Executing this selection corrects
input pressure to the Barometric
Standard value when the
pressure input is open to the
atmosphere.
The live value of the pressure
input is shown on this display so
the user can easily see the result
of the correction.
The upper pressure value for a
two-point field calibration within
the LRV to URV span.
Cal high point value is in inH2O
[4°C].
Range: LRV to URV
Default: URV
The lower pressure value for a
two-point field calibration within
the LRV to URV span. Cal low
point value is in inH2O [4°C].
Range: LRV to URV
Default: LRV
A ratio value that adjusts the
calibrated curve between the Cal
Low and High Points.
Error Adjustment Ratio
Range: 0.5 to 1.5
Default: 1.0
Press  to
enter menu
selection
Press  to
initiate action
Press  to
enter menu
selection
 Menu to
select
number.
 to enter
and shift to
the next digit
to the right
Press  to
enter menu
selection
 Menu to
select
number.
 to enter
and shift to
the next digit
to the right
Press  to
enter menu
selection
 Menu to
select
number.
 to enter
and shift to
the next digit
to the right
Revision 13

Advertisement

Table of Contents
loading

Table of Contents