Parts Required - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
8-13-2
Optional External Display Installation Instructions
8-13-2-1
Manpower
One person, Approximately 1 hour
8-13-2-2
Tools
Service Dongle
Standard Field Service Tool kit
Appropriate PPE
ESD
8-13-2-3

Parts Required

The LOGIQ E9 External Display Option kit is comprised of the following contents:
Remote Display plus adapter - H4910SM.
The Optional External Display offered with the LOGIQ E9 is the same as the one used in the
Voluson E8.
See: the installation instructions provided with the Kit (Document Number: KTD102148 Part Number:
KTI301000). It provides information on how to mount the additional Patient monitor to the wall and
connect it to the Voluson® system. The installation on the LOGIQ E9 is identical except for an adapter
needed to interface between the DVI-I output and the Monitor Cable.
On the LOGIQ E9, The VGA cable needs to be connected to the LOGIQ E9 rear panel using the DVI-I
to S-VGA adapter provided in the Kit. See:
Figure 8-72 S-VGA adapter between the LOGIQ E9 and External Monitor
8 - 310
. 2
EV
Section 8-13 - Option Installation/Replacement Instructions
Figure
8-72.
LOGIQ E9 S
M
ERVICE
ANUAL

Advertisement