Monitor Adjustment Procedure - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
6-2-3

Monitor Adjustment Procedure

Table 6-1
Room Condition
Monitor
Dark Room for Radiology/Cardiology
Dim Room for Radiology/Cardiology
Bright Room
Factory Settings
6-2-3-1
Brightness for 19 inch LCD Monitors
To reduce the brightness, press the left arrow button. A popup box with the current brightness
setting will be displayed on the screen.
To increase the brightness, press the right button.
Push the mode button once to remove the popup from the screen and save the current values.
6-2-3-2
Contrast for 19 inch LCD Monitors
Use the recommended Monitor settings. The contrast is adjusted by the LOGIQ E9 software, when
selecting scanning modes and doing scanning adjustments. Changing the monitor's contrast settings in
one mode will influence the picture quality in other modes.
6 - 4
. 2
EV
Recommended Monitor Settings
19 inch
85
90
100
100
Figure 6-3 19 inch Adjustment Buttons - Upgraded to R6
Section 6-2 - Monitor adjustments
Contrast
23 inch
OLED
100
40
100
65
100
100
100
40
LOGIQ E9 S
ERVICE
Brightness
19 inch
23 inch
35
40
55
70
100
100
60
70
M
ANUAL
OLED
50
50
50
50

Advertisement

Table of Contents
loading

Table of Contents