GE
D
5573152-100, R
IRECTION
1-4-4
Electrical safety
1-4-4-1
Safe practices
Follow these guidelines to minimize shock hazards whenever you are using the scanner:
•
The equipment chassis must be connected to an electrical ground.
•
The unit is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. A separate power outlet with a 20 amp circuit breaker for 120
VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or 15 amp circuit
breaker for 100 VAC for Japan.
•
The power outlet used for this equipment should not be shared with other types of equipment.
•
Both the system power cable and the power connector must meet international electrical standards.
Connecting a LOGIQ E9 scanner to the wrong voltage level will most likely destroy it.
WARNING
WARNING
1-4-4-2
Probes
Follow these guidelines before connecting a probe to the scanner:
•
Inspect the probe prior to each use for damage or degradation to the:
-
-
-
-
-
-
•
Do not use a damaged or defective probe.
•
Never immerse the probe connector or adapter into any liquid.
•
The system has more than one type of probe port. Use the appropriate probe port designed for the
probe you are connecting.
Section 1-5
Label locations
Refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 2, or the latest version
of the LOGIQ E9 Release Notes.
CONSOLE MODEL
1 - 20
. 2
EV
housing
cable strain relief
lens
seal
connector pins
locking mechanism
Table 1-8
Basic User Manual and Release Notes per LOGIQ E9 console
NUMBER
5205000
5205000-2, -3
5205000-4, -5
5205000-6, -7
5205000-8
5205000-9
Section 1-5 - Label locations
ENGLISH BASIC
USER MANUAL
DIRECTION NUMBER
5180374-100
5335626-100
5389558-100
5450756-100
5496408-100
5573149-100
LOGIQ E9 S
ENGLISH
RELEASE NOTES
DIRECTION NUMBER
5180358-100
5335634-100
5389562-100
5449984-100
5476411-100
5573151-100
M
ERVICE
ANUAL