After the Exam is Over

Endocavitary Probe Handling Precautions

CAUTION
3-4
If the sterilization solution comes out of the endocavitary probe,
please follow the cautions below.
Sterilant Exposure to Patient (e.g., Cidex)—Contact with a
sterilant to the patient's skin or mucous membrane may cause
an inflammation. If this happens, refer to the sterilant's
instruction manual.
Sterilant Exposure from Probe Handle to Patient (e.g.,
Cidex)—DO NOT allow the sterilant to contact the patient.
Only immerse the probe to its specified level. Ensure that no
solution has entered the probe's handle before scanning the
patient. If sterilant comes into contact with the patient, refer the
the sterilant's instruction manual.
Sterilant Exposure from Probe Connector to Patient (e.g.,
Cidex)—DO NOT allow the sterilant to contact the patient.
Only immerse the probe to its specified level. Ensure that no
solution has entered the probe's connector before scanning the
paitent. If sterilant comes into contact with the patient, refer the
the sterilant's instruction manual.
Endocavitary Probe Point of Contact—Refer to the
sterilant's instruction manual.
LOGIQ E9
Direction 5454884-100 English Rev. 1