Circuit breaker

NOTE:
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
The Circuit Breaker is located at the rear panel of the system.
On supplies main power to all internal systems. Off removes
main power from all internal systems. The circuit breaker
automatically shuts off power to the system in case of a power
overload.
If a power overload occurs:
1. Turn off all peripheral devices.
2. Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the On position; DO
NOT hold the switch in the On position. If the Circuit Breaker
switch remains On, follow the Power On procedure.
If the Circuit Breaker switch does not remain in the On position
or trips again:
1. Disconnect the Power Cable.
2. Call Service immediately.
DO NOT attempt to use the system.
Figure 1-14. Circuit Breaker (located at the rear of the system)