Scanning a New Patient (continued)
HINTS
CAUTION
NOTE:
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
10. Store the images/loops to the clipboard.
To store the still image, press Freeze and run the cineloop
using the Trackball. Select the frame and press P1 (or the
assigned Print key).
To store the cineloop, press Freeze and run the cineloop
using the Trackball. Select the start/end frame and run the
selected loop. Press P1 (or the assigned Print key).
When you press a Print key during scan (Live Store), the
LOGIQ E9 stores the cine for the specified length of time
noted in the Time Span(s) preset, found under Time-Based
Store in Utility -> Application -> Print Controls.
When you press a Print key during scan (Live Store) with
ECG, the LOGIQ E9 stores the cine for the specified
number of heart cycles noted in the Number of heart cycles
preset, found under ECG-Based Store in Utility ->
Application -> Print Controls.
If you set the Number of heart cycles for more than 1 heart
cycle in Utility -> Application -> Print Controls, the LOGIQ
E9 stores the cine for the length of time specified in "Time
Span(s) x Number of heart cycles".
Refer to Chapter 16 for more information on image
storage.
11. When you have completed the study, press End Exam. The
image management screen displays. Select the images (still
frame or cineloop) you want to store or select Select All to
store all images. Select Permanent Store to store the
images permanently.
After completing the measurement, verify that the
measurement result window is updated before you send or
save the image.
Return to the patient screen automatically from the scan
screen when you select OK from the "ID is not unique"
warning message.
Beginning an Exam
1-67