Other Maintenance

Manually Moving the Operator Panel Up/Down
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
The LOGIQ E9 is equipped with an automatic Operator Panel
up/down, left/right, forward/backward control. But if the system
is powered down when you need to lower the Operator Panel,
you can use the manual release handle located at the back of
the unit to manually move the Operator Panel up/down.
To move the Operator Panel up/down using the manual control,
1. Locate the manual release handle which extends through
the air vent at the back of the LOGIQ E9.
Figure 3-17. Manual release handle
2. Move the manual release handle to the left with one hand
while pushing up/down on the Operator Panel to position it.
For information on resetting the Park Lock, see 'Park Lock
Reset' on page 1-22.