Important Notices

NOTICE
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
Do not attempt to install the system alone. General Electric,
Affiliate, or Distributor Field Engineers and Application
Specialists will install and setup the system. See 'Contact
Information' on page 3-42 for more information.
This medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and
other institutions which are environmentally qualified. The use of
this equipment in an inappropriate environment may cause
some electronic interference to radios and televisions around
the equipment.
Ensure that the following is provided for the new system:
A separate power outlet with a 20 amp circuit breaker for
120 VAC for 120V area, 7.5 amp circuit breaker for
220-240VAC for 220/240V area or 15 amp circuit breaker for
100 VAC for Japan.
Take precautions to ensure that the console is protected
from electromagnetic interference.
Precautions include:
Operate the console at least 5 meters (15 feet) away
from motors, typewriters, elevators, and other sources
of strong electromagnetic radiation (non-medical grade
UPS must be at least 2 meters (6 feet) away from
console).
Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent
electromagnetic interference.
Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.
Console Overview
1-5