Honeywell SMV800 User Manual page 70

Multivariable transmitter modbus communication
Hide thumbs Also See for SMV800:
Table of Contents

Advertisement

Advance Flow Review:
This screen provides an overview of all the Flow related parameters that are configured in above
steps.
If the "Raw Flow" value given in review page is a valid number but the actual flow displayed in
"Flow Config" screen is NAN, then provide soft/master reset to device. Flow can be set to NAN if
any of the parameters like viscousity and density becomes invalid based on algorithm type during
run time configuration.
Page 58
SMV800 Modbus User's Guide
Revision 1.0