11.2.5.25 MV (Mitral Valve)
11.2.5.26 Manual Trace
Note
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
•
Pulmonary Veins
(chapter 'Pulmonary Veins' on page 11-79 )
•
PAP (Pulmonary Artery Pressure Measurement)
Pressure Measurement)'
•
HR (Heart Rate)
(chapter 'HR (Heart Rate)' on page 11-81 )
There are different methods to perform measurements and calculations of the Mitral Valve in
the Spectral-Doppler mode:
•
Manual Trace
(chapter 'Manual Trace' on page 11-75 )
•
To measure the Items One by One
11-76 )
1.
After obtaining a feasible Doppler spectrum, press the [Calc] key on the control panel.
2.
Select the measurement item [MV] and then select [Manual Trace]. A cursor appears on
the Doppler spectrum.
3.
Move the cursor to the start point of the period and press the right or left trackball key
[Set] to fix the marker. Begin trace.
To re-adjust the traced line, press the upper trackball key [Undo] repeatedly.
4.
Trace to the end of the period and press the [Set] key again to fix the mark.
Measurements and Patient Worksheets (Reports)
(chapter 'PAP (Pulmonary Artery
on page 11-80 )
(chapter 'To measure the Items One by One' on page
11-75