Volume Mode
9.10.2 After STIC Calculation
Note
9-112
After the calculation, the Estimated Heart Rate is shown on the menu area. The previously
selected format and the last acquired 4D sequence will be shown on the monitor.
Guidance and precautions for interpretation of STIC-images:
Judge if the shown estimated heart rate is reasonable. Visually rule out phase errors and
other failures of the acquisition before confirming the scan by selecting the [Accept] key.
Always adopt a critical attitude to images created in STIC mode.
Be aware that any diagnostic conclusion must not be drawn from STIC-images alone,
but have to be checked with other diagnostic procedures.
If you are in doubt about a structure seen in STIC mode, consult the original 2D-images
for clarification.
Please note that the accuracy of measurements in STIC-images is limited and can be
lower than measurements in B-images. Note for Users in Germany: Die Genauigkeit
kann die KBV-Richtlinien unterschreiten.
In case the acquisition fails, select the [Cancel] key and perform the
acquisition again. Selecting this key will bring you to the pre-acquisition mode.
Select the [Accept] key. The "Vol. Cine" menu in freeze mode appears on the menu area.
Upon [Accept] a certain number of volumes will be stored in the cine memory. The 4D
Sequence can be reviewed volume by volume.
on page 9-95
'Volume Cine'
The 3D menu appears.
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4