11.1.3.1 Generic Measurements
11.1.3.1.1 Distance 2 Points
11.1.3.1.2 Slope
Note
11.1.3.1.3 Time
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
This measures the vertical distance (tissue depth) between two points. The measurement
procedure is the same as the distance measurement in 2D Mode. review: Distance 2 Points
(chapter '2D Mode Measurements' on page 11-5 ) .
1.
To measure the time and slope, select the [Slope] item in the menu area. A cursor
appears on the screen.
2.
Move the cursor to the start point of the measurement and press the right or left trackball
key [Set] to fix the marker. A second cursor appears.
3.
Move this cursor to the second point of the measurement and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
1.
To measure a horizontal time interval, select the [Time] item in the menu area. A line
appears on the screen.
2.
Move the line to the start point of the measurement and press the right or left trackball
key [Set].
Measurements and Patient Worksheets (Reports)
11-15