8.7.12 Gently Colors
PW
-
8.7.13 Invert
PW
X
Forward flow indicates:
Reverse flow indicates:
8.7.14 Line Density
PW
-
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
CW
-
X
Gently Color defines the transition between color and gray scale information. With [Gently
Color] the embedding of the color into 2D mode is performed smoothly with less color flash.
Switch on/off the Gently Color function.
CW
-
X
This function inverts the spectrum display in relation to the direction of flow. The displayed
spectrum is inverted around the baseline. The velocity or frequency scale changes
accordingly. Use Invert when necessary to change the spectral display orientation. It is
possible in both freeze and scan mode.
[Invert] key on the control panel.
key unlit:
Normal:
key lit:
Inverted:
CW
-
X
This function determines the line density within the TD box. The lower the line density, the
greater the distance between lines and the size of the color pixels.
Select [-] or [+] on the [Line Dens] key and adjust the line density.
max. value: 10 min. value: 1 step size: 1
CF
PD
-
CF
PD
X
Blood flow towards the transducer (RED) (spectrum
above baseline)
Blood flow away from the transducer (BLUE) (spectrum
below baseline)
Forward flow above the baseline (RED) Reverse
flow below the baseline (BLUE)
Reverse flow above the baseline (BLUE)
Forward flow below the baseline (RED)
CF
PD
X
Doppler Modes
HD-Flow
TD
-
-
HD-Flow
TD
X
X-
HD-Flow
TD
X
X
8-37